Omega HH503 Instrukcja Użytkownika Strona 12

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 16
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 11
10
CALIBRATION PROCEDURE
Note: The following calibration procedures should be performed only by
qualified technicians who have access to the following:
Equipment: The class of calibration should be 10 times greater than the
measured meter.
Step 1: Short the J3 with jumper.
Step 2: Apply 65 mV to the T1 input from Volts Calibrator and wait till the
reading becomes stable, then press the SET key until the display shows 0.1.
Step 3: Apply Type K 0°C to the input T2 and wait till the reading
becomes stable, then press the SET key until the meter shut
s off by itself.
Step 4: Remove the jumper from J3.
Przeglądanie stron 11
1 2 ... 7 8 9 10 11 12 13 14 15 16

Komentarze do niniejszej Instrukcji

Brak uwag